Preparing The Immersion Fitting; Mounting The Sensor In The Immersion Fitting; Mounting The Sensor Cleaner (/Jc Option) - YOKOGAWA FD30 Instruction Manual

Table of Contents

Advertisement

7

5.2.2 Preparing the immersion fitting

Prepare the immersion fitting by unscrewing
the screw piece (4) from the bottom of the
fitting. Remove the stopper (7), which can
be discarded, and the conductive bushing
(5). The screw piece will be used to fix the
DO sensor in the fitting. When a flexible
conduit (protecting hose) will be used,
the pigtail part (2) can also be discarded.
When an immersion fitting with cleaner
is used the cleaning cage will be used to
fix the DO-sensor. In that case the screw
piece can also be discarded. In case of
a FD30V27-00 version the top piece and
the holder piece should be mounted on
the customers pipe using the supplied
adhesive. Ensure that the sealing will be
watertight.
5.2.3 Mounting the sensor in the
immersion fitting
Pull the sensor cable through the O-ring
into the fitting and through the pigtail (see
Figure 3). In case a flexible conduit will
be used please remove the pigtail and
follow the instructions delivered with the
protecting hose option. Push the O-ring
over the cable assembly until it lies on the
knurled ring. Plug the cable assembly into
Figure 2
the sensor body, if this was not performed
before. Secure the cable assembly with the
knurled ring. Place the sensor in the fitting
and put the bushing (essential for sensor
diagnostics) around the sensor as far as
it will go. Then screw the screw piece, or
the cleaning cage, in again. This will fix the
sensor in place by compressing the O-ring.
Tighten the pigtail at the top of the fitting.
IM 12J06K04-E-E

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pb30

Table of Contents