Asembling the Sensor in the Holder

For details on how to assmble the sensor in the holder or fitting, see the appropriate instruction
manual for each holder or fitting.
CAUTION
Exercise care not to contaminate or wet the tip of the sensor cable when assembling the sensor.
If you will not begin wiring the sensor cable immediately, take the protective measures necessary
to prevent the assembly from being damaged.
Sensor mounting in the case where flocs and other matter are contained in the SUM
Stopper
The stopper projecting from the membrane assembly locknut serves to prevent the membrane
from being in contact with the ground and thus being damaged. If there are flocs in the SUM, the
stopper may become clogged. Thus, assemble the sensor to the holder so that the stopper is
positioned downstream in the SUM.
3.3

Connecting the Sensor Cable

The cable of the dissolved oxygen sensor must be connected to the terminals of the dissolved
oxygen converter. For information on connections, refer to the following instruction manuals:
FLXA402 4-Wire Converter (IM 12A01F01-02EN), DO402G Dissolved Oxygen Converter
(IM 12J05D02-01E), FLXA202/FLXA21, DO202 2-wire Transmitter instruction manual (IM
12J05C01-01E/IM 12A01A02-01E) or WTB10-DO Relay Terminal Box (IM 12J05W02-01E).
Dissolved oxygen
converter terminals for
connection to sensor
cable
*1: Liquid earth to detect membrane failure.
This function is available when using
PB350G/PB360G float type holder.
Connect only when using this function.
Figure 3.5
Now the explanation on the installation and wiring is completed. For details on operation during
startup or for regular operation after the entire measuring system is brought into operation, refer
to the FLXA402 4-Wire Converter instruction manual (IM 12A01F01-01EN, etc.), the DO402G
Dissolved Oxygen Converter instruction manual (IM 12J05D02-01E), the FLXA202/FLXA21,
DO202 2-wire Transmitter instruction manual (IM 12J05C01-01E/IM 12A01A02-01E). The main-
tenance and service of the dissolved oxygen sensor is explained in Section 4 that follows.
<3. Installing the Sensor and Connecting the Cable>
F07-1.ai
Jumper
IE
21
17
13
13
Connection of Sensor Cable (example of DO402G)
T1
T2
SE
RE
S
11
12
14
15
16
11
12
14
15
16
(*1)
Sensor cable
F08.ai
IM 12J5B3-01E
6th Edition : Nov.20,2018-00