< 2. Installation >
Sensor Holder
Figure 2.6
To install the sensor cable in the sensor holder, first remove the protector screwed onto the
sensor holder end and then remove the protective foam piece (for shipping; thus, it is not
necessary after the sensor is installed in the holder). Pass the sensor cable through the
O-ring then attach the O-ring to the sensor flange (see Figure 2.7).
When passing the sensor cable through the holder, if the inside of the holder is dirty or wet,
take special care to keep the cable dry by covering the sensor cable end with a polyethylene
bag or the like. Pull out the sensor cable and KCI solution supply tube downward through
the holes of Waterproof Cap.(see Figure 2.7)
For details to install the sensor onto a holder, refer to relevant User's Manuals.
When Variopin connecter is attached to a sensor, conncect the sensor to SSA and WU11 cable
first, then insert the sensor into the holder pipe. For further information on sensor connection,
read IM 1206S01-00EN-(P), IM 1206S01-01Z1-(P).
CAUTION
DO NOT hold the lower part of the sensor when screwing it into a Variopin connecter. Otherwise
the excessive force may damage the sensor.
Clamp (Screw)
Cleaner Holder
F2.6.ai
Removal of Sensor Holder (with Cleaner)
< 2. Installation >
IM 12B7K1-02E
11th Edition : Dec.14,2018-00