Adding The Trigger Input Pod; Prerequisites; Installing The Trigger Input Pod - Agilent Technologies 81200 Installation Manual

System
Table of Contents

Advertisement

Adding the Trigger Input Pod

If you have ordered the 8-line Trigger Input Pod E4805A Option 002 separately, you need to
install it into the Agilent 81200 System.

Prerequisites

Prerequisites for this installation are:
• an Agilent E4805A Central Clock Module, the serial number of this module must be higher
• the Agilent E4873A User Software Revision 2.0 or higher
• a flat screw driver and a TORX No.362 screw driver

Installing the Trigger Input Pod

To connect the Trigger input pod, it is necessary to unplug the E4805A module from the
mainframe, to open the module and to plug the pod cable into the board.
1. Switch off the system and disconnect the power cord
2. Disconnect all cables from the E4805A/B Central Clock Module. If necessary, mark the
3. Remove the E4805A/B Central Clock Module from the mainframe
4. Unscrew the 7 screws to open the side panel of the module using a TORX screw driver
Figure 35
Open the Agilent E4805A/B Central Clock Module
than DE37800166
cables to make it easier to re-connect the cables into the right position
Unscrew the 7 screws
Modify Your System
Adding the Trigger Input Pod
The TORX screw driver
is required
59