Error
Description
Potential cause and remedy
«ERR 3»
No pressure
An adequate pressure cannot be gene-
in the cuff
17
rated in the cuff. A leak may have
occurred. Check that the cuff is cor-
rectly connected and is not too loose.
Replace the batteries if necessary.
Repeat the measurement.
«ERR 5»
Abnormal
The measuring signals are inaccurate
result
and no result can therefore be dis-
played. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high
pressure too
(over 300 mmHg) OR the pulse is too
high
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
«LO»
Pulse too
The pulse is too low (less than 40
low
beats per minute). Repeat the measu-
rement.*
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through
the information in «Section 1.» carefully.
16

10. Safety, Care, Accuracy Test and Disposal

Safety and protection

• This device may only be used for the purposes described
in this instruction. The manufacturer cannot be held liable
for damage caused by incorrect application.
• This device comprises sensitive components and must be
treated with caution. Observe the storage and operating con-
ditions described in the «Technical Specifications» section.
• Protect it from:
water and moisture
impact and dropping
contamination and dust
direct sunlight
heat and cold
• The cuffs are sensitive and must be handled with care.
• Only pump up the cuff once fitted.
• Do not use this device close to strong electromagnetic
fields such as mobile telephones or radio installations,
as they can impair its function. We recommend observing
a minimum distance of 1 metre. If you cannot maintain
the minimum distance please check the proper function
of the device before use.
• Do not use this device if you think it is damaged or notice
anything unusual.
• Never open this device.
• If the device is not going to be used for a prolonged period
the batteries should be removed.
• Read the additional safety information in the individual
sections of this instructions.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware
of the risk of strangulation in case this device is supplied
with cables or tubes.
®
aponorm