Changing the sheath filter
Introduction
When to change
the sheath filter
Replacing the sheath cubitainer (page 26)
Replacing the waste cubitainer (page 27)
Specifications and requirements (page 17)
Troubleshooting (page 44)
This topic describes how to change the sheath filter. The sheath
filter is connected in-line with the sheath line. It filters the sheath
fluid as it comes from the sheath container.
We recommend changing the sheath filter assembly every 3 to
6 months. Increased debris appearing in an FSC vs SSC plot can
indicate that the sheath filter needs to be replaced.
Chapter 4: Maintenance
39