28
BD FACSFlow Supply System User's Guide
Replacing the
waste cubitainer
Caution: Biohazard! All biological specimens and
materials coming in contact with them are considered
biohazards. Handle as if capable of transmitting
infection and dispose of with proper precautions in
accordance with federal, state, and local regulations.
Wear suitable protective clothing, eyewear, and gloves.
To replace the waste cubitainer:
1. Press the Alarm button to silence the alarm.
2. Press the STNDBY button on the cytometer.
3. Loosen the waste cap assembly and carefully remove the
sensor probe from the waste cubitainer.
4. Place a cap, set aside from a new sheath cubitainer installed in
the system, onto the waste cubitainer, and tighten it.