Flushing The Bd Lsrfortessa System - BD FACSFlow Supply System User Manual

Supply system for bd lsrfortessa cytometers
Table of Contents

Advertisement

30
BD FACSFlow Supply System User's Guide
More information

Flushing the BD LSRFortessa system

Introduction
About the stainless
steel sheath
containers
9. Pour 2 L of undiluted bleach into an empty cubitainer.
10. Apply a biohazard label to the top of the empty 20-L
cubitainer and label it as WASTE.
11. Load the empty cubitainer into the right side of the
BD FACSFlow supply system unit.
Make sure the biohazard and WASTE labels are facing the
front of the carrier and can be read.
12. Insert the sensor probe into the waste cubitainer, replace the
neck support collar, and tighten the cap assembly.
Caution! Replace and tighten all caps when
reassembling the containers and cubitainers. Loose
caps can result in fluid leakage.
13. Press Restart.
Flushing the BD LSRFortessa system (page 30)
Replacing the sheath cubitainer (page 26)
This topic describes the procedure for flushing the BD LSRFortessa
system. Flush the system every two weeks.
The cytometer system is shipped with a stainless steel container for
sheath fluid. This container is replaced by the sheath cubitainer
when the BD FACSFlow supply system is installed. Save the
stainless steel container and tubing and use it when flushing the
system.