Replacing the air filter
Introduction
Procedure
14. Reconnect the air and fluid lines coming from the sheath
plenum.
Flushing the BD LSRFortessa system (page 30)
This topic describes the procedure for replacing the air filters. An
air filter is located on the cap assemblies of the sheath and waste
cubitainers.
Caution: Biohazard! Treat a contaminated air filter as
biohazardous waste.
To replace an air filter:
1. Remove the air filter.
Hold the silicone tubing with one hand and pull off the filter
with the other hand. See the following figure.
Chapter 4: Maintenance
37