BD FACSFlow Supply System User's Guide
Procedure: Sheath
Empty alarm
More information
8. Press the Restart button on the control panel.
The indicator light switches off.
Pressing Restart triggers a check of the waste sensor in the
cubitainer to verify that the waste has been emptied. If the
waste was emptied, the indicator light switches off.
If the indicator light does not switch off, contact
BD Biosciences.
The Sheath Empty alarm is triggered when the sensor probe detects
that the sheath cubitainer is empty.
To respond to a Sheath Empty alarm:
1. Press the Alarm button on the control panel to silence the
alarm.
2. Check the sheath cubitainer to determine whether it is empty.
3. If the sheath cubitainer is empty, replace it.
4. Press the Restart button on the control panel.
Shutting down the system (page 22)
Replacing the sheath cubitainer (page 26)
Replacing the waste cubitainer (page 27)
Troubleshooting (page 44)