5. Remove the neck support collar and save it for the next
cubitainer.
Item
Description
1
Neck support color
2
Waste cubitainer
3
Sheath cubitainer
6. Place the 2-inch long waste sensor probe into the 500-mL
beaker containing the 10% bleach solution you have prepared.
Caution! Avoid splashing the 10% bleach solution
inside the rim of the cap assembly. Doing so can
saturate the air filter, which should then be
changed. See
instructions.
7. If necessary, label the cubitainer as WASTE and affix a
biohazard label onto the cubitainer.
8. Discard the waste according to federal, state, and local
regulations.
Replacing the air filter (page 37)
Chapter 4: Maintenance