Flushing the BD LSRFortessa system with a BD HTS

Introduction
13. Repeat steps
2
Solution Concentrate.
Note: The BD Detergent Solution Concentrate must be diluted
before use. Mix one full 15 mL bottle of BD Detergent
Solution Concentrate into 985 mL of DI water to make 1 L
total.
Caution! Do not mix BD Detergent Solution
Concentrate and bleach because they produce chlorine
gas.
14. Repeat steps
2
BD Detergent Solution Concentrate.
15. At the cytometer, disconnect the air and fluid lines coming
from the stainless steel container.
Note: Make sure to rinse any remaining bleach out of the
stainless steel container when the procedure is completed to
avoid corrosion.
16. Reconnect the air and fluid lines coming from the sheath
plenum.
Flushing the BD LSRFortessa system with a BD HTS (page 33)
This topic describes the procedure for flushing the BD LSRFortessa
system when the BD
installed. Perform the system flush every two weeks.
See the BD High Throughput Sampler User's Guide for additional
information on cleaning the HTS system.
through
11
with 1.5% dilution of BD Detergent
through
11
with DI water in place of
®
High Throughput Sampler (HTS) is
Chapter 4: Maintenance