Vital Signs Monitor Service Manual
3.3 Safety Requirements
1
SHOCK HAZARD- the power receptacle must be a three-wire grounded outlet. A
hospital grade outlet is required. Never adapt the three-prong plug from the monitor
to fit a two-slot outlet.
2
Do not touch signal input or output connector and the patient simultaneously.
3
Devices connecting with monitor should be equipotential.
4
Do not switch on the monitor until all cables have been properly connected and
verified.
3.4 Installing the Monitor
To install the monitor on a flat surface
Place the monitor on a flat surface. Make sure the surface does not vibrate, and is free of
corrosive medicine and dust.
CAUTION
Figure 3-1 iM3 on a flat surface
- 9 -
Installation