Vital Signs Monitor Service Manual
8.3.3 Replacing the Main Control Board
○
○
○
1
Screws
iM3 Wires linking main control
○
2
board and FCC rotary knob board
iM3 power linking wire of interface
○
3
board
iM3 Wires linking main control
○
4
board and FCC interface board
○
5
iM3 key board linking wire
Separate the front and
assembly
back
Assemble the main control board in the reversed order. Connect the wires and fix the main unit.
○
1
○
2
3
4
Disconnect the
related wires.
○
○
6
5
○
6
Lithium-manganese button battery
○
7
Micro SD card
○
8
LCD backlight linking wire interface
Touch screen linking wire
Alarm indicator linking wire
- 43 -
Disassembling the Monitor
○
8
○
7
Remove the screws securing
the main control board.