Vital Signs Monitor Service Manual
3 The service personnel must be familiar with the operation of this monitor. Refer to iM3
Vital Signs Monitor User Manual for details.
4 Perform periodic safety test to ensure patient safety. Safety tests should include
leakage current measurement and insulation testing. It is recommended to perform
the safety test every two years. You are responsible for any requirements specific to
your country.
5 Disconnect the monitor from power before replacing the fuses which are with the
identical specifications.
6 SHOCK HAZARD – Do not remove the top panel cover during operation or while
power is on. The unit cover must be removed only by authorized service personnel.
7 SHOCK HAZARD – Do not attempt to connect or disconnect the power cord with wet
hands. Make sure that your hands are clean and dry before touching the power cord.
8 Accessory equipment connected to the analog and digital interface must be certified
according to the respective IEC/ EN standards (e.g. IEC/ EN 60950 for data
processing equipment and IEC/ EN 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the valid version of the system standard IEC/ EN
60601-1.Anybody that connects additional equipment to the signal input connector or
signal output connector to configure a medical system must ensure that the system
complies with the requirements of the valid version of the system standard IEC/ EN
6060-1. If you have any question, please consult our technical service department or
your local distributor.
9 Do not directly solder the lead wire and the battery terminal.
10 The safe loads of the wall mounting bracket and the trolley are 7.5 kg and 11 kg
respectively. Exceeding the safe load may cause bracket to fail and the device to fall.
1 The device is designed for continuous operation. Avoid splashing water over the
device.
2 Do not operate the device when it is damp or wet. Avoid using the device immediately
after relocating it from a cold environment to a warm and humid environment. If the
monitor gets damp or liquid pours on the monitor, please contact the service
personnel of EDAN.
3 While the battery is charged, used or stored, keep it away from objects or materials
with static electric charges.
WARNING
CAUTION
- 5 -