Download Print this page

KYOWA WGA-910A Series Instruction Manual page 31

For cc-link option
Hide thumbs Also See for WGA-910A Series:

Advertisement

10-3. TO READ THE PRE-SET VALUES OF THE COMMANDS (BY USING THE REQUEST 2)
To read the pre-set values from the product, set the [WRITE/READ] to "ON." The product outputs the pre-set values.
Before reading the [Command data response], make sure the [Command processing response 2] is "ON."
(1) With the commands for the Measureing Condition (for details, see "10-1. TO WRITE THE PRE-SET VALUES
OF THE MEASUREING CONDITION (OPERATION PATTERN) LIMITED DATA (BY USING THE
REQUEST 1)"), enter the measureing condition in the [Measureing Condition] of the remote register M
You are not required to set the item with the commands for the general settings and commands for the control
input signals.
(2) Enter the commands in the [Command No.] of the remote register M
(3) Make sure the [WRITE/READ] of the remote output is ON.
(4) Change the [Command processing request 2] of the remote output. OFF [0]
(5) Make sure the change of the [Command processing response 2] of the remote input. OFF [0]
Then change the [Command processing request 2] of the remote output. ON [1]
(6) The product stores the response in the [WRITE/READ response] of the remote input and in the [Command No.
response], [Command data response], and [Measureing Condition response] of the remote register R
Check the contents of the [Command data response].
Command processing
request 2 (RY0002)
Command processing
response 2 (RX0002)
WRITE_READ
(RY0003)
Measureing Condition
(RWw000F)
Command No.
(RWw000E)
WRITE_READ response
(RX0003)
Command No. response
(RWr000E)
Command data response
(RWr000C to 000D)
Measureing Condition response
(RWr000F)
10-4. CPU NORMAL OPERATION
You are able to check whether the product is working properly by using the CPU normal operation.
The product turns ON and OFF the CPU normal operation every second when the product is working properly.
CPU normal operation
(RX0006)
ON [1]
OFF [1]
ON [1]
OFF [1]
ON [1]
OFF [1]
ON [1]
OFF [0]
Approx.1 sec. Approx. 1 sec.
29
R.
ON [1]
OFF [0]
(4)
(3)
(1)
(2)
R.
ON [2]
M.
(5)
(6)
(6)
(6)
(6)

Advertisement

loading

This manual is also suitable for:

Wga-910a-4