3.0 Setting up and starting up
Testing the deflection mechanism
• Slowly operate the deflection lever
• Check if full deflection is achieved (see chapter Technical Data).
Limitations in deflection can indicate an endoscope defect. To avoid serious damage to the endoscope, only use the endoscope
!
if the deflection works smoothly and without limitation.
Risk for patients
!
Risk for patients due to the use of products with sharp edges or damaged surfaces.
• Do not use these products.
Damaged products
!
Do not use products with damaged camera chip (e. g. recognizable by image interferences), damaged glass surfaces or
stubborn deposits, which cannot be removed by cleaning.
Send damaged products to the manufacturer or authorized repair centre. Authorized repair centres can be inquired from the
manufacturer.