EXTERNAL CALIBRATION (NOT AVAILABLE ON APPROVED MODELS) 
1) Press the [on/off] key to turn the power on.  
2) Press  the  [Smpl]  and  [Print]  keys  at  the  same  time  during  the  self‐checking 
test. 
3) The display will show unload. Remove any weight from the stainless steel pan.  
4) When the stable indicator shows press the [‐>0/T<‐] key.  
5) The  display  will  show  the  first  weight  that  you  can  use  to  calibrate  the  unit. 
You can change this value by pressing the [Mode] key. The weights that can 
be used are as follows:  
 
Model # 
Weight 1 
Weight 2 
Weight 3 
 
Model # 
Weight 1 
Weight 2 
Weight 3 
 
 
 
1) Once the calibration weight has been selected press the [‐>0/T<‐] key 
2) The  display  shows  LoAd.  Place  your  weight  on  the  pan.  Press  [‐>0/T<‐]  key 
once the stable sign is shown.  
3)  The display will show PASS. Remove the weight from the pan.  
 
NOTE: If the calibration fails retry.  The balance will show FAlL H (when the weight is 
higher)  or  FAlL L  (when  the  weight  is  lower).  Repeat  the  process  using  the  correct 
calibration weight. 
© Adam Equipment Company 2018
HCB 123 
HCB 153
60g 
50g
120g 
100g
‐ 
150g
HCB 1002  HCB 1502
500g 
500g
1000g 
1000g
‐ 
1500g
HCB 302
HCB 602
100g
200g
200g
400g
300g
600g
HCB 2202
HCB 3001
1000g
1000g
2000g
2000g
3000g
HCB 602H 
200g 
400g 
600g 
HCB 6001 
2000g 
4000g 
6000g 