Change Plate / Dish Type - Cantium Scientific MicroBio MB2 Operating Manual

Bioaerosol sampler
Hide thumbs Also See for MicroBio MB2:
Table of Contents

Advertisement

Change Plate / Dish Type

The MicroBio MB2 can accommodate both 55mm/65mm contact plates or
90mm petri dishes. The MicroBio MB2 is factory fitted with the petri dish
springs (Cantium Scientific Limited part number A-00070), but these can be
removed and replaced with the supplied contact plate springs (Cantium
Scientific Limited part number A-00068).
The springs for both types have
slots to allow a degree of adjustment
to suit dish/plate manufacturer
variations.
To change the spring type remove
the three screws as highlighted
using the 2mm hex-key supplied
with the kit. Lift off the support plate
and springs, then refit using the
same screws, support plate and the
required springs. Care must be
taken not to over-tighten or cross-
thread any of the screws.
Replacement screws are available from Cantium Scientific Limited, part
number A-00232 or contact your local distributor.
31