P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement
P0002W006!
2!
23 July 2010!
!
!
/ 1 12

Advertisement

loading

Summary of Contents for Cantium Scientific MicroBio MB2-RSH

  • Page 1 Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement MicroBio MB2-RSH! Installation and Operation Supplement! Document No.! P0002W006! Issue:!! Date:! ! 23 July 2010! Issue 2! / 1 12...
  • Page 2 Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement Issue 2! / 2 12...
  • Page 3 Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement Document Issue! Technical Specification! Installation! Fitting the Connector for Unterminated Cables! Connecting the Control Unit and remote Sampling Head! Calibration! Operation! Cleaning and bio-decontamination! Issue 2! / 3 12...
  • Page 4 Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement Issue 2! / 4 12...
  • Page 5: Document Issue

    Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement Document Issue! Issue Description Date Revision by Approved First release 20 June 2010 SNJP Added section on bio- 23 July 2010 SNJP decontamination Issue 2! / 5 12...
  • Page 6: Technical Specification

    Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement Technical Specification! Please refer to the MicroBio Manual P0001W003 for full detailed specification of the MicroBio MB2.! The MB2-RSH model is capable of driving a remote sampling fan up to a distance of 20m from the control unit.!
  • Page 7 Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement 2. Fit connector housing and strain relief onto cable end.
 3. Remove cable clamp.
 4. FIt frame onto cable end and refit cable clamp.
 Issue 2! / 7 12...
  • Page 8 Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement 5. Solder cable onto connector pins. The connector body has pin numbers embossed onto the surface adjacent to the pins.
 i. Pin 1: Red! ii. Pin 2: Blue! iii. Pin 3: Screen
...
  • Page 9 Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement 7. Slide the assembly, strain relief and metal shell together.
 
 8. Align shell so that the brass thread of the frame body can be seen, then replace the housing screw and tighten.
...
  • Page 10: Connecting The Control Unit And Remote Sampling Head

    Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement Connecting the Control Unit and remote Sampling Head! Now position the remote sampling head and connect the fly lead on this to the extension cable end labelled “RSH”. The connectors have a screw locking mechanism. Screw the two halves together by twisting the outer ring of the plug from the the remote sampling head connector into the cable socket on the extension cable.
  • Page 11: Calibration

    Cantium Scientific Limited! P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement Calibration! The MB2-RSH can be calibrated in the same manner as any other MicroBio bioaerosol sampler.! Remove the remote sampling head from its normal location, disconnect the extension cable from the control unit and then plug the remote sampling head directly into the control...
  • Page 12: Operation

    30 minutes! ! Aeration time! ! 150 minutes! As the MicroBio MB2-RSH has a fan within the remote head it is recommended that during the decontamination process the MicroBio MB2-RSH is set to sample 2000 litres several times so that the H vapour may be drawn through the head, fan and exhaust for optimum decontamination.!