Cantium Scientific MicroBio MB2 Operating Manual page 23

Bioaerosol sampler
Hide thumbs Also See for MicroBio MB2:
Table of Contents

Advertisement

Finally, define the time delay before sampling starts. Adjust from 0 for no
delay up to 3 hours using
Set Delay
01:00:00
The MicroBio MB2 will calculate the volume each sequence is to take and
the interval time between each sample.
Sequential Sampling process will commence, displaying a countdown if a
delay was set. The first sample will be taken, then wait for the calculated
interval before taking the next sample. This will continue until the total
volume has been sampled over the required time.
During sequential sampling, the display will show a count of sequences (top-
left), the total volume taken (bottom-left), time to next sample (top-right) and
time remaining until sequential sampling is complete (bottom-right).
1/5
200L
At the end of sequential sampling, the screen will indicate sampling is
complete and the total sampled volume. The MicroBio MB2 will automatically
switch off.
At any time, either during set-up or sampling, the process can be terminated
by holding down MENU. If sampling is interrupted, the unit will display a
summary of what has been sampled.
or
01:00:00
09:00:00
then press START.
22