During Recording
value is ignored when determining alarm trigger conditions, minimum/maximum values
and other statistical calculations.
The Paused Readings feature is useful, for example, when you need to temporarily
remove the probe from the monitored location to inspect goods, but you do not
wish to trigger an alarm as a result of you handling the probe.
It allows the logger to acclimatize to the environment
again, before further readings are processed.
After a button press the display shows
the
symbol.
PAUSED
ignored reading has been recorded.
The option is set in the
configuration with LogTag
number of readings after the last button press.
Paused readings are specially marked in the graph and data
listings.
How long
PAUSED
button. It will show longer, if the button is pressed just after the logger takes a
reading, but shorter if you press the button just before. For example, if you
configure a logging interval of 10 minutes and 2 paused readings, the time
is shown could be as short as 10 minutes, but as long as 20 minutes.
Probe is disconnected
The display shows
temperature probe is disconnected. This function is useful if the probe is permanently
mounted, but you wish to take the TREL30-16 to a computer for downloading its data.
Any values recorded during this time will not be
taken into account when calculating statistics values
or alarms.
If, for example, the temperature was above 0 °C for
12 hours (which would normally trigger an alarm
event), but the probe was disconnected for 3 hours
during this time, no alarm would be generated.
TREL30-16 User Guide
PAUSED
will turn off as soon as the last
Advanced Settings tab
®
Analyzer and is expressed in
is displayed depends on when between readings you press the
instead of a temperature value if the logger's remote
next to
during
26
PAUSED