Marking A Reading With An Inspection Mark - LogTag Recorders TREL30-16 Product User Manual

Ultra-low temperature recorder with display
Table of Contents

Advertisement

During Recording
At 1:49am on the next day the display shows the following:
 
The temperature has returned to within the accepted range (none of the alarm
l
threshold markers are visible), but the alarm remains present, as it has not been
inspected and cleared.
 
The day summary has shifted by 1 day as the display time has passed through
l
midnight (00:00).

Marking a reading with an inspection mark

When you press the
reading taken will be identified in the downloaded data with an inspection mark.
If the
Allow stopping with the Stop button
registered when you press the
stopping the logger. An inspection mark will also be recorded if you clear an alarm.
TREL30-16 User Guide
button while the TREL30-16 is recording, the next
REVIEW
/MARK
REVIEW
/MARK
feature is enabled, a mark will also be
button, but do not complete the process of