Extended Specifications on G1316B SL/G1316C SL Plus

Extended Specifications on G1316B SL/G1316C SL Plus

The 1200 series G1316B SL/G1316C SL Plus thermostatted column
compartment is usable from 10 °C below ambient up to 80 °C for flow ranges
up to 5 ml/min and up to 100 °C for flow ranges up to 2.5 ml/min. Additional
heating and cooling devices are available for the G1316B SL/G1316C SL Plus
to reduce the risk of additional dispersion at low flow rates see
page 30. These devices can be installed in any position in the column
compartment, see
SL)"
Figure 8
The standard column compartment is equipped with a 3 µl and 6 µl heater or
cooler. Both can be set at the same or different temperature. To reduce the
delay volume, a kit
page 104) has been set up for installing a small heaters with 1.6 µl internal
delay volume and also a new cooling device with 1.5 µl internal volume is
available.
30
"Installation of Heater and Cooling Devices (G1316B
on page 47.
New additional heater and cooling devices
("G1316B SL/G1316C SL Plus Capillary System Kit"
Figure 8
on
1200 Series TCC User Manual
on