Replacing the Column Compartment's Firmware

The installation of older firmware might be necessary:
• to keep all systems on the same (validated) revision, or
• if third part control software requires a special version.
To upgrade/downgrade the TCC's firmware the following steps have to be
performed:
When
If new version solves problems of currently installed version or after exchange of the TCC main board
(CCM) the version on board is older than previous installed one.
Tools required
LAN/RS-232 Firmware Update Tool, or Instant Pilot G4208A or Control Module G1323B
Parts required
Description
Firmware, tools and documentation from Agilent web site
Preparations
Read update documentation provided with the Firmware Update Tool.
1 Download the module's firmware, the LAN/RS-232 FW Update Tool Version
2 Load the firmware into the TCC as described in the documentation.
The G1316B SL requires firmware revision A.06.02 or higher (main and resident).
N O T E
The G1316C SL Plus requires firmware revision A.06.10 or higher (main and resident).
1200 Series TCC User Manual
2.10 or above and the documentation from the Agilent web
http://www.chem.agilent.com/scripts/cag_firmware.asp.

Replacing the Column Compartment's Firmware

7
Maintenance
85