4
How to optimize the Column Compartment
Optimizing the Performance of your Column Compartment
Optimizing the Performance of your Column Compartment
For best performance results of the column compartment:
• Use short connection capillaries and place them close to the heat
• Use the left heat exchanger for small volume columns, for example, 2–3 mm
• For even lower band-broadening, the heat exchanger can be by-passed and
• Keep the left and right heat exchanger temperature the same unless you do
• Assure that the front cover is always closed.
52
exchanger. This will reduce heat dissipation and external band-broadening.
i.d. columns at flow rates of less than 200 µl/min.
the column is placed well between the heat exchanger fins.
specific applications.
1200 Series TCC User Manual