Installing the Column Compartment

Installing the Column Compartment

Installing the Column Compartment
Parts required
#
1
1
Preparations
Locate bench space.
Provide power connections.
Unpack the Column compartment.
Module is partially energized when switched off, as long as the power cord is
WA R N I N G
plugged in.
Risk of stroke and other personal injury. Repair work at the module can lead to
personal injuries, e. g. shock hazard, when the module cover is opened and the
instrument is connected to power.
Valve properties are read from the valve head RFID tag during initialization of the
C A U T I O N
module. Valve properties will not be updated, if the valve head is replaced while the
module is on.
Selection of valve port positions can fail, if the instrument does not know the
properties of the installed valve.
38
Description
Column compartment
Power cord
For other cables see text below
Never perform any adjustment, maintenance or repair of the module with the top
cover removed and with the power cord plugged in.
The security lever at the power input socket prevents that the module cover is taken
off when line power is still connected. Never plug the power line back in when cover
is removed.
Always switch off the instrument when replacing the valve head.
1200 Series TCC User Manual