Ceiling Suspended X-Ray Tube Support; General; Configuration; Weight - Canon OMNERA 400A Operator's Manual

Auto positioning digital radiographic system
Table of Contents

Advertisement

Technical Specifications

Ceiling Suspended X-ray Tube Support

8.2 Ceiling Suspended X-ray Tube Support

8.2.1 General

Rotation range ceiling (beta)
Rotation range tube arm (alpha)
Column (Z stroke)

8.2.2 Configuration

OTC

8.2.3 Weight

OTC
Tube and collimator
Ceiling wagon
Column
Ceiling rail Y (4 m standard)

8.2.4 Electrical Characteristics

Mains voltage

8.2.5 Classification

Classification according to IEC 60601-1.
Class
Applied part
Protection against ingress of water
Mode of operation
Use of anaesthetic mixtures
OMNERA®
>340°
>±135°
1750 mm
The OTC is a mechanical part of an X-ray
system.
Maximum 240 kg
40 kg maximum allowed weight
95 kg
40 kg
28 kg
230 VAC, 50/60 Hz center tapped single
phase 4 A
Class I equipment. All dead metal parts of
the equipment are electrical connected to
protective earth.
Type B
IPXO
Intermittent operation: 20%, maximum 1
min. ON / 4 min. OFF
The equipment is not suitable for use in the
presence of flammable anaesthetic mix-
tures with air or with oxygen or with nitrous
oxide.
0073-095-418-EN Rev. 5.1
Page 126
Operator´s Manual
2020-09-25

Advertisement