Wallstand Examination Procedures; Wall Flexible Modes - Canon OMNERA 400A Operator's Manual

Auto positioning digital radiographic system
Table of Contents

Advertisement

Operating the System

Wallstand Examination Procedures

4.8 Wallstand Examination Procedures

4.8.1 Wall Flexible Modes

4.8.1.1 General
The Wall Flexible modes are designed for examinations of patients standing up against a
vertical moving imaging unit.
The movement up/down of the OTC, will only change the tube position, the Wallstand will not
follow.
4.8.1.2 Wall Flexible Movements
There is 3 different Wall Flexible Modes, all accessible from the Arcoma Service program,
Adjust WallFlexible parameters. The Wall Flexible Movements shall be set at the installation,
and will thereafter be valid for all Auto positions using the Wall Flexible mode.
• Not Blocked — All OTC movements are allowed.
• Beta blocked and Sideways supervised — Blocked beta movement, movement will turn
the servo off. It is possible to move a distance of half the detector length (landscape), be-
fore turning off the servo.
• Beta and Sideways blocked — It is only possible to move the OTC upwards and down-
wards. Blocked beta movement, movement will turn servo off. Blocked non-FFD direction-
al (X/Y) movement, movement will turn servo off and block exposure.
OMNERA®
Fig. 4-15 Adjust WallFlexible parameters.
0073-095-418-EN Rev. 5.1
Page 94
Operator´s Manual
2020-09-25