Operating the System
Detector, Wallstand

4.9.1 14x17 Detector, Wallstand

4.9.1.1 Method to Load the 14x17 Detector, Wallstand
The method of setting the detector to the detector holder is as follows:
1. Pull the detector tray toward you until it locks.
Note!
• Install the detector with the detector tray pulled into the locked position. When pulling the
detector tray, the button on the side of the tray will first recede before snapping back into
its original position when it locks.
2. Insert the detector into the detector holder, as shown below and set it by pushing it in, un-
til it clicks.
CAUTION!
• Confirm that the latch is going up firmly, as shown below.
3. While pressing and holding the button of the detector tray, return it back to the inside of
the detector holder.