Operating the System
Detector, Wallstand

4.9.2 17x17 Detector, Wallstand

4.9.2.1 Method to load the 17x17 detector
Note!
• Depending on left- or right-operated wallstand, the location of the detector tray and posi-
tion of button and latches is different.
The method of setting the detector to the detector holder is as follows;
The following instruction describes the detector operated from the left side.
1. Pull the detector tray toward you. Make sure the detector tray is completely brought out.
2. Insert the detector into the detector tray as shown below and set it by pushing the detec-
tor, holding down the latch.
3. While pressing and holding the button of the detector tray, return it back to the inside of
the detector holder.