OAKWORKS CFPMFXH User Manual page 37

Imaging tables
Hide thumbs Also See for CFPMFXH:
Table of Contents

Advertisement

SPECIFICATIONS
PRODUCT SPECIFICATIONS
Weight
Shipping Weight
Lifting Capacity
Materials of Construction
Aluminum Filtration
Equivalence and thickness
of Carbon Fiber Tabletop
ENVIRONMENTAL CONDITIONS
Conditions
Normal Use
Storage & Transport
ELECTRICAL SPECIFICATIONS
Designed for:
Input Service (All electric
tables)
Current Draw
Maximum Momentary Cur-
rent Consumption
Voltage Output to Actuators
Electric Shock Protection
Tabletop Applied Part
Ingress Protection Rating
Mode of Operation
NORTH AMERICA
359-452 lbs.
452-544 lbs.
500 lbs. (750 lbs. Bariatric)
ALUMINUM EQUIVALENCE
Non-Bariatric Tops
Bariatric Tops
All Extensions
1" (2.5 cm.) Table Pad
2" (5 cm.) Table Pad
This product complies with United States Department of Health and Human Services radiation performance standards,
21 CFR Sub chapter J, in effect at the time of manufacture for radiographic tables.
Temperature
50° (10°C) to 104° (40°C)
-20° (-29°C) to 135° (57°C)
CFPM100 Tables Only
NORTH
AMERICA
120 VAC/15 amp/60
Hz
5.8 amps
9.0 amps
120 VDC
Intermittent Operation: MAX 2 minutes ON / 18 minutes off
This product contains no latex.
100 kVp, HVL of 3.6mm = 1.10 (1.35 max)
100 kVp, HVL of 3.6mm = 1.50 (1.79 max)
100 kVp, HVL of 3.6mm = 1.66 (2.00 max)
100 kVp, HVL of 3.6mm = .88 (1.07 max)
100 kVp, HVL of 3.6mm = .88 (1.07 max)
Humidity
20% to 60% RH
20% to 95% RH
CFPM400, CFPM401, CFPM301,
CFLU401, CFUR401 & CFUR301
EUROPE
NORTH
AMERICA
230 VAC/10
120 VAC/15 amp/60
amp/50/60 Hz
3.0 amps
8.2 amps
4.5 amps
12.3 amps
220 VDC
120V Column
Class 1 Equipment
Type B Applied Part
IPX0
EUROPE
163-205 kg.
205-247 kg.
227 kg. (340 kg. Bariatric)
Atmospheric Pressure
98 to 105 kPa
98 to 105 kPa
Tables
EUROPE
220 VAC/10
Hz
amp/50/60 Hz
4.5 amps
6.7 amps
24V top,
24V top,
220V Column
THICKNESS
1.65" (4.2 cm)
2.18" (5.5 cm)
.5" (1.3 cm)
1" (2.5 cm)
2" (5 cm)