Alarms
The new alarm limit is now active. The cursor returns to
the
symbol.
Adapting alarm limits
Volume
When ventilation settings have been made, Primus
Infinity Empowered can automatically adapt the alarm
limits for minute volume MV and the airway pressure
P
to the current parameters in Volume Mode, Vol-
AW
ume AF, Pressure Mode, and Pressure Support (op-
tional).
Press the soft key auto-set limits (A).
The alarm limits menu opens automatically.
The alarm limits for MV and P
lighted by dark green background (B).
To quit the alarm limits menu:
Push the rotary knob or the
The new alarm limits for MV are calculated by Primus
Infinity Empowered from the measured value for the
minute volume MV in Volume Mode, Volume AF, Pres-
sure Mode, and Pressure Support (optional).
Setting the alarm tone
The volume of the alarm tone can be set in the configu-
ration menu.
24
A
B
MV
6.0
3.6 8.4
P
25
10 30
AW
are adapted and high-
AW
key.
Quick Reference Guide Primus Infinity Empowered SW 4.5n
Alarms in the Standby mode
All technical alarms, e. g. failure of equipment compo-
nents and a number of special operating states, are also
indicated by Primus Infinity Empowered when in Stand-
by.
A corresponding message appears in the alarms field,
but without any acoustic warning.