GRAPHTEC MT100 User Manual page 79

Mountcorder
Hide thumbs Also See for MT100:
Table of Contents

Advertisement

Settings and Measurement
(9)-1 OTHR settings
LCD brightness
Screen Saver
Power On Start
Room Temp.
Temp. Unit
Burn Out
AC Line cycle
Date/Time
Language
Return to default settings Returns all the settings to factory default.
3-40
Setting items
Sets the brightness of the LCD backlight.
Available selections are Light, Medium, and Dark.
Automatically turns off the display if the MT100 is not operated within a specified interval.
Available selections are Off, 10, 30 s, 1, 2, 5, 10, 30, 60 min.
Turning off the display frequently using the Screen Saver function allows longer lifetime of the
LCD screen.
Sets the feature which initiates measurement as soon as the MT100 is turned on.
Disable
Enable
This parameter enables room temperature compensation settings when thermocouples are used.
Internal
External
Toggles the temperature unit between °C and °F.
°C
°F
Sets a feature which checks sensor burnout in a thermocouple.
Off
On
Selects the frequency of the AC line used.
50Hz
60Hz
This parameter sets the date and time.
Date/Time Sets the internal clock of the device.
Internet
Time
Sets the MT100's display language.
Available selections are Japanese, English (US), English (UK), French, German, Chinese,
and Korean.
Description
Disables the Power On Start function.
Enables the Power On Start function.
The MT100's room temperature compensation settings are used
(usually, use this parameter).
This parameter is set to enable room temperature compensation settings in external devices.
Celsius
Fahrenheit (the scaling function is compulsorily enabled)
Disables burnout check.
Enables periodic burnout check.
CAUTION
During a burnout check, voltage is applied to the MT100.† Therefore, set Burn Out to Off
when MT100 is connected inparallel with other devices to avoid any effect from these voltages.
For areas using line frequency of 50 Hz.
For areas using line frequency of 60 Hz.
CAUTION
This setting specifies the frequency in which noise can be eliminated with the digital filter
function.
Note that the digital filter is enabled when the sampling frequency is set to 500 ms or above.
Enter the correct date and time.
This setting is used to adjust the internal clock via the network.
*For details, see the next page.