GRAPHTEC MT100 User Manual page 62

Mountcorder
Hide thumbs Also See for MT100:
Table of Contents

Advertisement

(2)-2 Backup Settings
MT100 provides a back up function where captured data can be backed up periodically (see figure below).
This section allows you to set data back up conditions.
Measurement input
Setting items
Backup Intervals
Backup Desination
Folder Name
(2)-3 Statistical Calculation
Two calculations can be performed in MT100.
This section allows you to set details of the statistical calculation.
Setting items
Calc. Settings 1 (2)
CHECKPOINT
• Operation results are displayed in the Digital + Calculation Display screen.
• Calculation will start upon power ON. Pressing the Start key to begin measurement will clear the calculation.
Automatic back up to FTP server
MT100
Internal memory
Automatic back up  
to USB memory
Description
Sets the interval to perform back up of captured data.
Off, 1, 2, 6, 12, 24 h
Sets the location where back up data will be stored.†
USB1
Data is backedup in the USB memory.
This is enabled only when data is captured in the internal Flash memory.
FTP
Data is backed up in the FTP server in the network.
* FTP settings in the FILE menu is required (see 3-32).
Specifies the folder where data will be saved.
ex) \GRAPHTEC\TEST\20071204
Description
off
Calculation is not performed.
Average
Displays the simple average value during data capture.
Max
Displays the maximum value during data capture.
Min
Displays the minimum value during data capture.
Peak
Displays the peak value data during data capture.
RMS
Displays the effective value of the data during data capture.
The following formula is used.
ΣD
R.M.S
D : data
n : number of data
*
FTP server
2
/n
Settings and Measurement
Ethernet
USB memory
3-23