Download Print this page

Overheating Mode - Otto Bock Genium 3B1-3 User Instructions

Hide thumbs Also See for Genium 3B1-3:

Advertisement

User Instructions

Overheating mode

Typically caused by heavy use of the knee joint (e.g. long
downhill walking, biking in basic mode)
Overheating is indicated by long vibrations every 5 seconds
Reaction of the knee joint depends on the temperature
increase:
Stance phase flexion resistance increases (with
increasing temperature)
Swing phase flexion resistance decreases (with
increasing temperature)
Consequence → User has to walk slowly
Reduce activity until the hydraulics have cooled down
OPG, PreFlex, sitting and stance function are still
available