mA Source Verification
1. Turn ON the 719 Calibrator.
When the display indicates a pressure and current value, for example:
0.00 PSI
0.000mA
2. Push three times, the lower display should read:
Source 4.000mA
The lower display will flash 4.000mA / OL until the circuit is complete.
3. Connect a test lead from Red Calibrator mA terminal to the Fluke 8508A I+ terminal.
4. Connect a test lead from the I-terminal of the Fluke 8508A to the Calibrator COM
terminal.
5. Set the 8508A to DC mA.
6. Change the output mA value by pressing (Steps are 4 mA, 8 mA, 12 mA, 16 mA,
20 mA), or scroll the mA output value between 0...24 mA using or .
7. Verify that the current readings on the Fluke 8508 are within the limits shown in
Table 70.
8. Disconnect the test leads and turn off the 719 Calibrator.
Fluke 719
4.0000 mA
12.0000 mA
24.0000 mA
Table 70. 719 mA Source Verification
Fluke 8508A
3.9974 mA to 4.0026 mA
11.9962 mA to 12.0038 mA
23.9944 mA to 24.0066 mA
Process Calibrators
Verification
55