719 Calibration (V1.0); Ma Measure - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

719 Calibration (V1.0)

mA Measure

1. Hold down both and on the 719 Pressure Calibrator. Push and release .
2. The display shows CAL xxx, as long as and are pressed. Where xxx is a
number, indicating the number of times the 719 has been through the calibration
adjustment. When and are released, PASS is displayed.
3. A password is required to enter calibration mode. The password is 917. Enter the
most significant digit first. The password appears below PASS. Use  to increase the
displayed number,  to decrease it. When the required digit appears, push  to
proceed to the next digit, or the password check. If the password entry was correct,
the 719 will go to the CAL mode. If the password is incorrect, BAD PASS shows on
the display and the 719 returns to the normal operation mode.
4. In the CAL mode the upper display shows an uncalibrated reading. This reading is
approximately ±10 % of full scale of the applied reading. When  is pressed, the
calibration value is calculated relative to this reading.
5. The lower display indicates what function and point is being calibrated.
6. In mA Measurement calibration mode, the 719 should display:
mA on the upper display and
CAL 0 mA on the lower display.
7. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on the
719 Calibrator (black to black and red to red).
8. Set the Fluke 5520A to 0.00000 mA.
9. When 0.0 mA is sourced from the 5520A the upper display indicates approximately
0.000 mA. When  is pressed, the unit establishes the calibration constant for that
point and the lower display changes to CAL 12 mA. The upper display still reflects
the applied current so it does not change until after 12 mA is applied.
10. Set the Fluke 5520A to 12.0000 mA. The upper display reads approximately
12.000 mA.
11. Push  on the 719 Calibrator.
12. When 12.000 mA is sourced from the 5520A, the upper display indicates
approximately 24.000 mA. When  is pressed, the unit establishes the calibration
constant for that point and the lower display changes to CAL 24 mA. The upper
display still reflects the applied current so it does not change until after 24 mA is
applied.
13. Set the Fluke 5520A to 24.0000 mA.
The upper display reads approximately 24.000 mA.
14. Push  on the 719 Calibrator.
CHEC appears on the lower display.
15. The CHEC mode allows for a quick check of the calibration constants that have just
been stored. Changes to the applied current are reflected in the display.
16. Push  to continue. The upper and lower display changes to - - - - -. The Source,
mA, Ref appear on the LCD, and the unit will proceed enters the mA output
calibration section.
Process Calibrators
Calibration
81

Advertisement

Table of Contents
loading

Table of Contents