Fluke 71X Series Calibration Manual page 27

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

1
Range
-1 to 1
-12 to 15
PSI
1.0000
15.000
bar
0.0689
1.0342
mbar
68.948
1034.2
kPa
6.8948
103.42
kg/cm2
0.0703
1.0546
cmH20@
70.309
1054.6
4 °C
cmH20@
70.434
1056.5
20 °C
inH20@
27.681
415.12
4 °C
inH20@
27.730
415.95
20 °C
mmHg@
51.715
776.25
0 °C
Overload Protection
Fuseless overvoltage protection
Temperature Coefficient
0.005 % of range per °C for temperature ranges of between −10 °C to 18 °C and 28 °C to
55 °C.
Loop Supply
24 V dc nominal
Pressure Module Input
Determined by pressure module
Table 22. 717 Range and Resolution
15
30
100
-12 to 30
-12 to
100
30.000
100.00
2.0684
6.8947
2068.4
6894.8
206.84
689.48
2.1092
7.0307
2109.3
7030.9
2113.0
7043.4
830.42
2768.1
831.89
2773.0
1551.5
5171.5
300
500
1000
-12 to 300
0 to
0 to
500
1000
300.00
500.00
1000.0
20.684
34.474
68.947
20684
34474
68948
2068.4
3447.4
6894.8
21.092
35.153
70.307
21093
35154
70309
21130
35217
70434
8304.2
13840
27681
8318.9
13865
27730
15515
25858
51715
Process Calibrators
Specifications
1500
3000
5000
0 to
0 to
0 to 5000
1500
3000
1500.0
3000.0
5000.0
103.42
206.84
344.74
NA
NA
NA
10342
20684
34474
105.46
210.92
351.53
NA
NA
NA
NA
NA
NA
41521
83042
NA
41595
83189
NA
NA
NA
NA
10000
0 to
10000
10000
689.48
NA
68948
703.06
NA
NA
NA
NA
NA
17

Advertisement

Table of Contents
loading

Table of Contents