4. Set the Fluke 5520A to 0.00000 mA and push .
5. When 0.0 mA is sourced from the 5520A the upper display indicates about
6. Set the Fluke 5520A to 24.0000 mA.
7. The upper display reads approximately 24.000 mA. Push  on the 718 Calibrator.
8. The CHEC mode allows for a quick check of the calibration constants that have just

Pressure Measure

The 718 Pressure Calibrator has built-in temperature compensation. Instruments being
calibrated should be in a stable temperature environment for several minutes before
calibration. Calibration facilities should be maintained near 23 °C nominal.
Re-calibration (re-characterization) is performed in terms of 'PSI' pressure units. Inputs
using other pressure units (ie. kPa and bar) must be mathematically converted.
For versions 3.00 and later, has changed to .
1. Remove two-wire test lead from the 718 Calibrator and carefully attach the pressure
The use of TEFLON tape at the pressure fittings strengthens the seal.
2. The lower display shows rAnGE. This is an indication to choose the correct model of
For unit with V2.0 or less, the upper display shows 30, which is the default pressure
range. Pressing
to 30 PSI. For a≥ version 3.0 push to uprange, hold to down range. Match the
displayed range to the measurement range of the 718.
3. After selecting the proper range, push
4. Vent the system to ambient air to ensure 0.00 PSI.
5. The lower display indicates C 0 PSI and the upper display indicates the current
76
"0.000 mA". When  is pushed, the unit establishes the calibration constant for that
point and the lower display changes to "CAL 24 mA". The upper display still reflects
the applied current so it does not change until after step 6 when 24 mA is applied."
CHEC appears on the lower display.
been stored. Changes to the applied current are reflected in the display. Push  to
continue. The lower display changes to:
rAnGE
fitting of the deadweight tester to the pressure jack on the 718 Calibrator.
the 718 Calibrator. Choose the correct range. The appropriate range for the 718 is as
follows:
718 1G (1)
718 30G (30)
718 100G (100)
718 300G (300)
changes the sensor range to 100 PSI, pressing
reading. When the unit is vented (step 5) the upper display will read 0.00 ±
approximately 10 % of full scale. When
the next pressure to apply. Apply the requested pressure that shows on the lower
display and when the upper reading is stable, push
appears on the display.
Note
Note
on the 718Ex Calibrator.
is pressed, the lower display indicates
. Repeat this until CHEC
changes it back