714 Specifications - Fluke 71 Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71 Series:
Table of Contents

Advertisement

Temperature Coefficient

714 Specifications

Firmware Earlier than v2.0
TC Type
J, K, T, E, L, U
B, R, S
Maximum input voltage: 30 V
Temperature Coefficient: 0.05 x specified accuracy per °C for temperature ranges – 10 °C to 18 °C and
28 °C to 55 °C
Maximum input voltage: 30 V
1.800.561.8187
0.005 % of ohms range per °C for temperature ranges –10 °C to 18 °C and 28 °C to
55 °C. Ohms ranges are 400 Ω, 1.5 kΩ, and 4.0 kΩ.
714 Calibrator specifications vary based on the version of the instrument. To display the
firmware version for your instrument, start with the 714 off, push and hold , then push
the power button. Find the section heading below for the displayed version and use the
specification tables in that section to test and calibrate the instrument.
Table 11. 714 Temperature Measure and Thermocouple Simulate
Resolution
0.1 °C
1 °C
Table 12. 714 Millivolt Measure and Source
Range
-10 mV to 75 mV
www.
Error
±(0.3 °C + 10 μV)
±(0.3 °C + 10 μV)
Resolution
±(0.025 % of range (75 mV) + 1 count)
0.01 mV
.com
Process Calibrators
Reference Junction Error
±0.2 °C
±0.2 °C
Accuracy
information@itm.com
Specifications
11

Hide quick links: