Accuracy test

We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). The verification time
period is 2 years. Verification is carried out in accordance with the
document Р 1323565.2.001-2018 «State system for ensuring the
uniformity of measurements. Non-invasive blood pressure meters.
Verification Method» Confirmation of initial verification, an elec-
tronic version of the verification certificate, is on the website of the
Federal Information Fund for Ensuring Uniformity of Measure-
ments - www.fgis.gost.ru
Information on verification is on the company website -
www.microlife.ru

Disposal

Batteries and electronic devices must be disposed of in

accordance with the locally applicable regulations, not with

domestic waste.

10. Guarantee

This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
 Batteries and parts that become worn with use are not included.
 Opening or altering the device invalidates the guarantee.
 The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
 The cuff has a functional guarantee (bladder tightness) for 2
years.
Please contact your local Microlife-Service (see foreword).
BP B2 Basic

11. Technical Specifications

Operating condi-
10 - 40 °C / 50 - 104 °F
tions:
15 - 95 % relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
277 g (including batteries)
Dimensions:
131 x 90 x 60.5 mm
Measuring proce-
oscillometric, corresponding to Korotkoff
dure:
method: Phase I systolic, Phase V
diastolic
Measurement range: 20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % des Messwertes
Spannungsquelle:
4 x 1.5 V alkaline batteries; size AA
Mains adapter DC 6V, 600 mA (optional)
Battery lifetime:
approx. 920 measurements
(using new batteries)
IP Class:
IP20
Reference to
EN 1060-1 /-3 /-4; IEC 60601-1;
standards:
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Date of production: first four digits of the serial number of the
device. First and second digit – the week of the year / third and
fourth digit – the year of production.