Assembly Of The Welding Adapter With Flange Or Threaded Connector; Welding Adapter With Flange Connector - ABB SensyMaster FMT230 Commissioning Instruction

Thermal mass flowmeter
Hide thumbs Also See for SensyMaster FMT230:
Table of Contents

Advertisement

FMT230, FMT250 THERMAL MASS FLOWMETER | CI/FMT230/250-EN REV. B

Assembly of the welding adapter with flange or threaded connector

Welding adapter with flange connector

Dimensions in mm (in)
1 Centering pin
2 Groove for O-ring
Figure 15: Dimensions in mm (in)
h – sensor length
263 (10.35)
425 (16.73)
775 (30.51)
*
The limitation of the maximum pipe diameter only applies for installations with a sensor element in the middle of the pipe. In case of larger or non-round
cross-sections, a non-centered position of the measuring element in the piping is considered in the calibration.
3 Connection flange DN 25 (1 in)
4 Flow direction
Ø D – outer pipe diameter (min. / max.)
100 to 350 (3.94 to 13.78)
> 350 to 700 (> 13.78 to 27.56)
> 700 to 1400 (> 27.56 to 55.12)*