Changing The Type Of Protection - ABB SensyMaster FMT230 Commissioning Instruction

Thermal mass flowmeter
Hide thumbs Also See for SensyMaster FMT230:
Table of Contents

Advertisement

FMT230, FMT250 THERMAL MASS FLOWMETER | CI/FMT230/250-EN REV. B

Changing the type of protection

If you are installing in Zone 1 / Div. 1, the Modbus interface and digital outputs of models FMT230/250 can be operated with different
types of protection:
Modbus interface and digital output in intrinsically safe ia / IS design
Modbus interface and digital output in non-intrinsically safe design
If a device that is already operational is operated with a different type of protection, the following measures must be
implemented/insulation checks performed in accordance with applicable standards.
Original installation
New installation
Zone 1 / Div. 1:
Zone 1 / Div. 1:
Modbus interface and digital
Modbus interface and digital
outputs in non-intrinsically safe
outputs in intrinsically safe
design
ia / IS design
Zone 1 / Div. 1:
Zone 1 / Div. 1:
Modbus interface and digital
Modbus interface and digital
outputs in intrinsically safe
outputs in non-intrinsically safe
ia(ib) / IS design
design
Note
For further details on explosion protection, types of protection and device models, refer to the installation diagram in the annex!
Necessary test steps
500 × 1.414 = 710 V DC/1min
Test between terminals A / B, 41 / 42 as well as 51 / 52 and the terminals A, B, 41, 42, 51 and
the housing. When this test is performed, no voltage flashover is permitted in or on the
device.
Optical evaluation particularly of the electronic circuit boards, no visible damage or evidence
of explosion.
Visual inspection, no damage visible on the threads (cover, ½ in NPT cable glands).