Pipe Component And Weld-On Adapter With Ball Valve - ABB FMT400-VTS Operating Instruction

Thermal mass flowmeter
Hide thumbs Also See for FMT400-VTS:
Table of Contents

Advertisement

4.6.2
The ball valve version allows you to mount and dismount the transducer even with low overpressure in the pipe
letting very little gas escape.
The weld-on adapter is mounted as described in chapter 4.6.1.
NOTICE
When welding, it must be ensured that the seals in the ball valve do not overheat.
!
In there is any doubt, the ball valve must be removed before starting the welding.
When mounting and dismounting the transducer always make sure that the gas pressure inside the pipe is not
as high that the transducer is pressed out of the pipe component or weld-on-adapter. We recommend to use the
integrated hot tap fitting for gas pressures above 1.1 bar abs.
Fully open the ball valve before mounting the transducer. Fit in the appropriate gasket, thoroughly center the
transducer and fasten it.
Prior to dismounting the transducer always check that the pipe is depressurized or that the pressure in the pipe
is low enough to ensure safe dismounting. Only then undo the 4 screws on the DN 25 flange, dismount the trans-
ducer and close the ball valve.
NOTICE
Do not close the ball valve before removing the transducer. Otherwise. the ball may considerably dam-
!
age the cage or the sensor elements.
Weld-on adapter with ball valve for FMT400-VTS (Sensyflow VT-S)
(1)
(2)
(1) Centering pin on outlet side
(2) ball valve DN 40
D Pipe diameter (outside)
Fig. 4-3
42/14-36 EN
Thermal Mass Flowmeter FMT400-VTS, FMT400-VTCS (Sensyflow VT-S/VT-CS)

Pipe component and weld-on adapter with ball valve

Sealing ring groove
∅ d
min. 28 mm
Direction of flow
Installation of the weld-on adapter with ball valve
Mounting and installation
Weld-on adapter
(upon delivery)
Connection flange DN 25
(1)
(2)
∅ 48.3
Z-20233
Required accuracy of mounting
Centric mounting < ± 2 mm
Twist < ± 2°
Length h
Min./max.
of the
outer pipe
transducer
diameter
(in mm)
(in mm)
263
100...150
425
> 150...500
*
775
> 500...1150
*
This maximum pipe diameter
specification is only valid when
installing the sensor centrically in
the pipe. For larger diameters or
angular ducts a non-centric sen-
sor position is taken into account
for calibration.
17