Changing the type of protection

Depending on the model, the device may be designed to be installed in one of the applications listed. If you plan to use a device
that was installed in one type of protection rating in a different type of protection rating, some measures must be taken before
connecting voltage to the device.
Measures for devices with alarm signaling unit FAM540-B/C/D:
Original installation
XP or Ex d
U
= 60 V
M
IS or Ex ia
NI or Ex nA
U
= 60 V
M
1) Possible only if the maximum signal levels of U
Measures for devices with transmitter with or without LCD display FAM540-E/F.
Original installation
XP or Ex d
U
= 60 V
M
IS or Ex ia
NI or Ex nA
U
= 60 V
M
1) Possible only if the maximum signal levels of U
Change from one to two columns
New installation
1)
IS or Ex ia
— 500 V AC / 1min test between terminals 51 / 52 and 41 / 42 and terminals
— Visual inspection: no explosion, no damage.
NI or Ex nA
— 500 V AC / 1min test between terminals 51 / 52 and 41 / 42 and terminals
— Visual inspection: no explosion, no damage.
XP or Ex d
— Visual inspection: no damage to threads (cover, surface, 1/2" NPT cable input), cable
NI or Ex nA
No special measures required.
IS or Ex ia
1)
— 500 V AC / 1min test between terminals 51 / 52 and 41 / 42 and terminals
— Visual inspection: no explosion, no damage.
XP or Ex d
— Visual inspection: no damage to threads (cover, surface, 1/2" NPT cable input), cable
≤ 60 V (e.g., PELV or SELV circuits) were not previously exceeded.
M
New installation
IS or Ex ia
1)
— 500 V AC / 1min test between terminals 31 / 32 and 41 / 42 and terminals 31 / 32 /
— Visual inspection: No damage, especially to electronics boards.
— Visual inspection: no explosion, no damage.
NI or Ex nA
— 500 V AC / 1min test between terminals 31 / 32 and 41 / 42 and terminals 31 / 32 /
— Visual inspection: No damage, especially to electronics boards.
— Visual inspection: no explosion, no damage.
XP or Ex d
— Visual inspection: no damage to threads (cover, surface, 1/2" NPT cable input), cable
NI or Ex nA
No special measures required.
1)
IS or Ex ia
— 500 V AC / 1min test between terminals 31 / 32 and 41 / 42 and terminals 31 / 32 /
— Visual inspection: No damage, especially to electronics boards.
XP or Ex d
— Visual inspection: no damage to threads (cover, surface, 1/2" NPT cable input), cable
≤ 60 V (e.g., PELV or SELV circuits) were not previously exceeded.
M
Necessary test steps
51 / 52 / 41 / 42 and the housing.
51 / 52 / 41 / 42 and the housing.
gland, glass, housing, locking device for cover, suitable cable, etc.
51 / 52 / 41 / 42 and the housing.
gland, glass, housing, locking device for cover, suitable cable, etc.
Necessary test steps
41 / 42 and the housing.
41 / 42 and the housing.
gland, glass, housing, locking device for cover, suitable cable, etc.
41 / 42 and the housing.
gland, glass, housing, locking device for cover, suitable cable, etc.