CareFusion Alaris GP Directions For Use Manual page 27

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Large Volume Pump Configuration Settings
AIL LIMIT
BOLUS MODE
BOLUS RATE DEFAULT
BOLUS RATE MAX
BOLUS VOLUME MAX
OCCLUSION ALARM PRESSURE
PRESSURE MAX
INFUSION RATE MAX
KVO RATE
NEAR END OF INFUSION POINT
PRIMARY VTBI MAX
PRIME RATE
PRIME VOLUME MAX
SECONDARY INFUSION
SEC. INFUSION RATE MAX
SEC. VTBI MAX
The single bubble Air-in-line alarm setting.
Controls whether or not the pump allows bolus delivery method.
The default value for bolus rates.
The maximum permissible bolus rate.
The maximum permissible bolus volume.
The default occlusion limit.
The maximum permitted occlusion limit.
The maximum permissible infusion rate.
The maximum infusion rate when performing KVO delivery.
The point defined in terms of remaining time at which the infusion is deemed to be Near end i.e.
nearly complete.
The maximum VTBI for primary infusions.
The rate at which priming is performed.
Determines the maximum volume that can be purged as part of a given prime operation.
Allows the use of a secondary infusion (Piggyback) in the same channel.
The maximum permissible infusion rate for secondary infusions.
The maximum permissible setting for the Volume To Be Infused for secondary infusions.
1000DF00546 Issue 5
Alaris™ GP (Guardrails™) Volumetric Pump
Pump Configuration available via the Alaris Editor Software
26/50