Drug Library available via the Alaris Editor Software

The following drug parameters are only configurable via the Alaris Editor Software (PC based), see Alaris Editor help files for further
details.
Concentration Settings
CONCENTRATION UNITS
CONCENTRATION MIN
CONCENTRATION MAX
Dose Rate Settings
DOSE RATE UNITS
WEIGHT BASED UNITS
DOSE RATE DEFAULT
DOSE RATE SOFT MIN
DOSE RATE SOFT MAX
DOSE RATE HARD MAX
Bolus Settings
BOLUS MODE
BOLUS DOSE UNITS
WEIGHT BASED UNITS
BOLUS DOSE DEFAULT
BOLUS DOSE SOFT MIN
BOLUS DOSE SOFT MAX
BOLUS DOSE HARD MAX
BOLUS RATE DEFAULT
Pressure Settings
OCCLUSION ALARM LEVEL
Display of Units
Units are selected via the Alaris Editor.
Micrograms can be displayed as mcg or µg depending upon the configuration in the Alaris Editor.
Units can be displayed as U or units depending upon the configuration in the Alaris Editor.
The unit for concentration parameters.
The weakest permissible concentration for this drug.
The strongest permissible concentration for this drug.
The unit for dose rate parameters.
Controls whether weight based units are enabled or disabled for use.
The default dose rate for infusing this drug.
(For Alaris GP Guardrails Volumetric Pump only) The minimum permissible dose rate which does
not generate an alert on the pump.
(For Alaris GP Guardrails Volumetric Pump only) The maximum permissible dose rate which does
not generate an alert on the pump.
The maximum permissible dose rate for infusing this drug.
Controls bolus delivery method. These settings override the pump configuration settings in the
profile.
The unit for bolus dose parameters. Applies to Hands-on and Hands-free bolus.
Controls whether weight based units are enabled or disabled for use.
The default bolus dose for this drug. Applies to Hands-free bolus.
(For Alaris GP Guardrails Volumetric Pump only) The minimum bolus dose that the pump allows
before the user is required to confirm the selected bolus dose in response to a warning of the
possibility of under-infusion. Applies to Hands-free bolus.
(For Alaris GP Guardrails Volumetric Pump only) The maximum bolus dose that the pump allows
before the user is required to confirm the selected bolus dose in response to a warning of the
possibility of over-infusion. Applies to Hands-free bolus.
The maximum permissible bolus dose for this drug. Applies to Hands-on and Hands-free bolus.
The default bolus rate for this drug.
The occlusion alarm level can be set from L0-L8.
1000DF00546 Issue 5
Alaris™ GP (Guardrails™) Volumetric Pump
Drug Library available via the Alaris Editor Software