CareFusion Alaris GP Directions For Use Manual page 21

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Rate Lock (If enabled)
If Rate Lock is enabled, when the infusion rate has been set and the infusion started (or following a bolus infusion) the rate lock prompt
will appear on the main display.
To select the rate lock function press the YES softkey. Press the NO softkey if the rate lock is not required.
When rate lock is enabled, the following are unavailable:
• Changing the infusion rate / titration
• Bolus / Prime
• Switching the pump off
• VTBI over time infusions.
• Secondary infusions (if enabled)
To turn rate lock off:
d
1. Press the
button to access the options menu.
2. Select UNLOCK RATE and press the OK softkey.
To turn rate lock on:
d
1. Press the
button to access the options menu.
2. Select RATE LOCK and press the OK softkey.
Adjusting Existing Dosing or Protocol Infusions - Set By ml/h / Set by Doserate
To set doserate or flowrate in precise increments it may be necessary to switch between the rate adjust options SET BY DOSERATE and
SET BY ml/h. An arrow to the left of the rate display shows the rate changed when the
the infusion rate.
To set a doserate precisely the arrow must be pointing to the doserate (for example: mg/kg/h); the flowrate will be calculated from the
doserate.
To precisely set a flowrate the arrow must be pointing to flowrate (ml/h); the doserate will be calculated from the flowrate.
Selecting the SET BY ml/h Option
d
1. Press the
button to access the options menu.
2. Select the SET BY ml/h option using the
flowrate option, the arrow on the display will automatically select the flowrate, the flowrate can be adjusted if necessary.
Selecting the SET BY DOSERATE Option
d
1. Press the
button to access the options menu.
2. Select the SET BY DOSERATE option using the
the set by doserate option, the arrow on the display will automatically select the doserate, the doserate can be adjusted if necessary.
Dosing Summary
To review currently selected dosing information:
d
1. Press the
button to first access the options menu.
2. Select DOSING SUMMARY option using the
3. Review the information and then press the QUIT softkey.
Add Drug (Only available when infusing)
d
1. Press the
button to access the options menu.
2. Select ADD DRUG option using the
3. Select from DRUGS (A-Z) from the displayed list using the
4. Select drug name using the
Infusion Setup
To change the Infusion Setup, refer to 'Basic Features - Drugs and Dosing, Selecting the INFUSION SETUP' section.
f
keys and press the OK softkey indicated on the screen. This will select the set by
f
f
f
keys and press the OK softkey.
f
keys, press OK to confirm and then follow the on screen prompts as required.
1000DF00546 Issue 5
Alaris™ GP (Guardrails™) Volumetric Pump
f
keys and press the OK softkey indicated on the screen. This will select
keys and press the OK softkey.
f
keys, press OK to confirm.
20/50
Basic Features
keys are used to increase/decrease