Display
FLOW ERROR
FLOW ERROR (In secondary
infusion mode only)
FREE FLOW
BATTERY EMPTY
SAFETY CLAMP
SET MISLOAD
FLOW SENSOR DISCONNECT
WRONG SET
DOOR CLOSE INCOMPLETE
DO NOT USE
LEVER OPEN
Infusion
Status
Infusion stopped
Infusion stopped
Infusion stopped
Infusion stopped
Pump on hold
Pump on hold
Infusion stopped
Pump on hold
Pump on hold
Pump on hold /
infusion stopped
Infusion stopped
1000DF00546 Issue 5
Alaris™ GP (Guardrails™) Volumetric Pump
Cause
Action
Gross difference
between detected drops
and expected amount of
drops.
Unexpected drops
detected.
Uncontrolled flow
possible.
The internal battery is
exhausted. The pump
will automatically switch
off in the immediate
future.
Safety clamp broken or
missing.
Set loaded incorrectly.
Flow sensor unplugged
during infusion.
Safety clamp not
detected.
Safety clamp in non-
occluded position with
door open or obstructed.
Internal error has
occurred.
Door lever is open
29/50
• Clamp infusion set using roller clamp.
• Check flow sensor.
• Check fluid level in drip chamber.
• Hang secondary container above
primary.
• Check drops are from secondary
container when infusing.
• Flow sensor disconnection is
recommended.
• Clamp infusion set using roller clamp.
• Remove pump from use.
• Connect to power supply
immediately or switch pump off.
• Clamp infusion set using roller clamp.
• Replace infusion set.
• Investigate and correct set loading.
• Clamp infusion set using roller clamp.
• Investigate and correct set loading.
• Check / replace flow sensor or set VTBI.
• Clamp infusion set using roller clamp.
• Check set and close door.
• Replace infusion set (If necessary).
• Clamp infusion set using roller clamp.
• Investigate and correct set loading.
• Close door.
• Remove pump from use.
• Check door lever.
• Check lever hooks.
• Check lever is not obstructed, if so, free
obstruction.